Operation Manual

110
Cleaning the blood glucose monitoring system
11
kim_manual_0780
6
Leave the cassette compartment
cover open and close the tip
cover.
4
Check whether you can see any
dirt inside the tip of the cassette.
If so, remove this carefully.
If there is dirt or blood on the
guidance tabs, carefully dab the
guidance tabs with a slightly
dampened cotton swab.
If the measuring optics are
visibly soiled, clean the inside of
the meter.
kim_manual_0765
5
Carefully dab the measuring
optics and the front opening of
the meter with a slightly
dampened cotton swab.
H
Make sure that no liquid enters the meter. Avoid scratching the measuring optics.
man_08252963001_02_EN-AUS.indb 110 27.02.2018 10:12:43