Operation Manual

99
9
Checking the meter
You can also check yourself whether the control result is within the right concentration range.
To do so, compare the control result with the concentration table on the test cassette box.
The control result must be within the printed concentration range 2.
Note: The values shown in the illustrations are only examples.
1 = Unit of measurement, 2 = concentration range, 3 = control solution used
Reasons for Control test not OK
If Control test not OK Repeat control test is displayed, the control result is not within the concentration
range displayed. The following overview can help you to eliminate the cause of this.
If none of the causes apply, contact the customer support and service centre.
W
If Control test not OK Repeat control test is displayed, you can no longer guarantee that the meter and
test cassette are functioning properly. Blood glucose tests may then deliver incorrect test results.
Incorrect test results may cause the wrong therapy recommendation to be made and thus produce
serious adverse health effects.
Cause Action
The control solution was applied too early.
Only apply the control solution to the test area
when Apply drop is displayed.
The control solution was applied too late.
Apply the control solution to the test area
immediately after you have opened the applicator.
The wrong colour was chosen.
In the Control test menu, select the colour of the
text shown on the applicator label.
The control solution has expired.
Perform the control test only with control solution
that is not past its use by date. You can find the
use by date on the bottom of the applicator next
to the
symbol.