- ACCU-CHEK user guide Blood Glucose Monitor insulin pump

5. Press d to move to the TBR DURATION screen.
Press
a or s to correct and set the duration.
6. Press f to save and exit.
The new Temporary Basal Rate is immediately activated.
Note:
At any function screen you have three options to exit:
if you want to save the changes
Press f. Your pump returns to the RUN screen.
if you want to
undo the changes
Wait for your pump to return to the RUN screen (timeout), or
Press
d+a simultaneously (exit feature) to exit the function screen.
Set volume to 100% and press f.
Note:
Press d to loop between programming the basal rate percentage
and duration.
Note:
When you increase (or decrease) the Temporary Basal Rate the duration of
your previous Temporary Basal Rate increase (or decrease) appears. When
you use your pump for the first time, a default value appears.
Chapter 4: Boluses and TBRs ACCU-CHEK Spirit Insulin Pump
95