User guide
Chapter6:TechnicalInformation 82 ACCU-CHEK Customer Care Service Center 1-800-858-8072
Chapter6:TechnicalInformation
Product Limitations
Please read the literature packaged with your test drum to find the latest information on product specifications
and limitations.
Specifications
Display Organic Light Emitting Diode (OLED) display
Automaticpower-off After 60 seconds / 5 minutes (depending on testing status)
Powersupply AAA alkaline batteries (do not use rechargeable)
Batterylife 1,000 tests
Interface Infrared interface
LED/IRED Class I
Metertype Reflectance photometric
Measuringrange 10–600 mg/dL
Samplesize Refer to the test drum package insert
Measuringtime 5 seconds
Systemoperatingconditions 50 °F to 104 °F
Meterstorageconditions With batteries and test drum: 36 °F to 86 °F
Testdrumstorageconditions 36 °F to 86 °F
Relativehumidityoperatingrange 15 %–85 % relative humidity
Memorycapacity 500 blood glucose values with time and date
Dimensions 4.9 x 2.5 x 1.3 inches (121 x 64 x 32 mm) with lancing device
4.8 x 2.2 x 1.3 inches (121 x 57 x 32 mm) without lancing device
Weight
Approximately 5.2 ounces (147 g) with lancing device, batteries and test strip drum
Lancingdevicedepths 0.80, 0.95, 1.10, 1.25, 1.40, 1.55, 1.70, 1.85, 2.00, 2.15, 2.30 mm