Checking your meter
The test strip draws up the control solution through the black notch.
Move the test strip away from the control solution as soon as the
hourglass symbol
is displayed and the meter beeps (provided the
beep tone is turned on).
The test strip has then drawn up enough control solution, and testing
starts. The hourglass symbol
indicates that the meter is busy per-
forming a test. The meter automatically detects that the test strip has
drawn up control solution. The bottle symbol
appears additionally in
the display when the test starts.
Testing is complete after approx. 5 seconds. The result is displayed and
the meter beeps (provided the beep tone is turned on). The meter at the
same time saves the result together with the bottle symbol
. The saved
result is thus recognizable as being a performance check result.