Operation Manual

Table Of Contents
Control Tests
28
4
Touch the drop to the
front edge of the yellow
window of the test strip
until you see flash.
Do not put control
solution on top of the
test strip.
5
Remove the cap from the
control solution bottle.
Wipe the tip of the bottle
with a tissue.
Squeeze the bottle until a
tiny drop forms at the tip.
4
Insert the test strip into
the meter in the direction
of the arrows.
Place the meter on a flat
surface.
Check the useby date
on the test strip
container. Do not use
test strips past the
useby date.
1
Performing a Control Test
You need the meter with pre-inserted activation chip, a test strip, and control solution Level1, Level2, or
both. The control level is printed on the bottle label.
Select the control
solution to test. You will
enter the level later in
the test.
32
or
6
When you see flash,
there is enough control
solution in the test strip.
Wipe the tip of the bottle
with a tissue. Cap the
bottle tightly.