Operation Manual
31
2
2.4 Updating the Diary with Bolus Information
When the pump is synchronized with the meter, bolus information from the pump is imported into the meter.
For any manual pump bolus record in the meter database with a time stamp that is older than 10 minutes
and is not confirmed by the pump, the meter does away with the bolus record by changing the bolus amount
to “0.0” on the next meter and pump synchronization. This eliminates the possibility of having a bolus on
the meter that was not delivered on the pump.
If you used the bolus advice feature and delivered a bolus manually on the pump, it is important you update
the information in the diary to reflect the amount of carbohydrates eaten with the bolus to ensure accurate
bolus advice recommendations. If a bolus is delivered using an insulin pen or syringe (not through bolus
advice), it is important to enter this information in the meter diary to ensure accurate bolus advice
recommendations.
For instructions on how to add and edit data in your diary, see the Standard Owner’s Booklet.
51639_00048010156_C.indb 31 6/11/13 8:47 AM