Operation Manual
45
씰 Apply the drop of blood on to the middle of the orange-coloured,
square application area (pad). It is OK to touch this area when
applying blood, but do not attempt to spread the blood. If you
suspect the amount of blood may not suffice, you may apply a
second drop within 5 seconds.
씰 Your Accu-Chek Active meter beeps briefly (provided the beep
tone is enabled) to acknowledge application of blood and to
announce the start of testing. Make sure you do not move the
test strip during testing, or an incorrect result may be obtained.
45
man_04493168001_02_EnUK.indd 45 1/12/2006, 11:29:01 ðì