Operation Manual

Wipe the measurement optics components with a soft lint-free
cloth and/or a cotton swab. The cloth/cotton swab may be
slightly moistened with water or 70% alcohol. Make sure that no
liquid enters the meter itself. Avoid scratching the measurement
optics, as this will impair the measuring function.
When all of the components are thoroughly dry, you may slide
the test strip guide back onto the meter. Ensure it clicks back
into place.
Then perform a quality control test (see Chapter 6, “Checking
Accu-Chek Active”).