Operation Manual

101
Clean the measurement optics carefully with nothing other
than cold water, soft lint-free cloths and cotton swabs. For
disinfection you may use 70 % alcohol. Any other cleaning
agents may damage the meter or impair its measuring function.
Slide off the test strip guide towards you (see illustration).
After removal of the test strip guide from the meter, clean it with
cold water.
Afterwards you may wipe the test strip guide with 70% alcohol.
Wipe off the alcohol immediately from the test strip guide and
allow it to dry thoroughly.