User`s manual

40
For persons with vision loss – testing blood glucose
F
a
r
be
: P 2
8
1 – Pr
ü
fmi
tte
ln
u
mm
e
r: 2
05
2
The voice unit issues a message, “Drum partly used, perform quality control”, drawing your attention to the fact. The message
is announced following a test, together with the test result.
Perform a quality control check (see Chapter 20).
If the test produces a correct result, you may continue to use the test strips. If the result is incorrect, you need to insert a new
drum.
Following the quality control check, the bottle symbol disappears from the display. The drum symbol and the clock (where
appropriate) continue to flash in order to draw your attention to the fact that the meter cannot calculate the 90-day use-by
period or that the drum has been in use for more than 90 days. If you do not perform a quality control check, the flashing
bottle symbol appears with every test until you insert a new drum. The drum symbol and clock (where appropriate) only disap-
pear when you insert a new drum. All results are saved with the flashing drum symbol and, where appropriate, with the flash-
ing clock symbol.
i
If you insert a partly used drum in the meter, a dot appears in the small control window at the back. When you next
turn the meter on, it rotates the drum until it finds the first chamber containing a test strip. The fewer test strips there
are in the drum, the longer the process takes. The control window allows you to see the drum being rotated from one
chamber to the next.