User`s manual
237
For sighted persons: Using the meter and cleaning the devices
6
Safety class III
LED/IRED Class 1
Electromagnetic compatibility
This device meets the electromagnetic immunity requirements as per ISO 15197 Annex A. The
chosen basis for the immunity tests (for electrostatic discharge) was basic standard IEC 61000-
4-2. In addition, it meets the electromagnetic emissions requirements as per EN 61326. Its
electromagnetic emission is thus low. Interference from other electrically powered equipment is
not expected.
Performance analysis The performance data for the Accu-Chek Compact Plus system (Accu-Chek Compact Plus
meter with Accu-Chek Compact test strips) were obtained using capillary blood from diabetic
patients (method comparison, accuracy), venous blood (repeatability) and control solution
(reproducibility). The system is calibrated with venous blood containing various levels of glu-
cose. The reference values are obtained using the hexokinase method. For method comparison,
the results were compared with results obtained using the hexokinase method with deproteini-
zation (automatic analyzer). The hexokinase method is traceable to an NIST standard.