6
25 Technical data
25.1 Accu-Chek Compact Plus meter
Device type Accu-Chek Compact Plus
Catalogue No./Serial No See type plate on the back of the meter
Test principle Determination of glucose in fresh capillary blood by reflectance photometry. When using
different specimen material please refer to the package insert that came with the
Accu-Chek Compact test strips.
Blood glucose concentrations may be measured in whole blood or in plasma. Although you
always apply whole blood to the test strip, your meter displays blood glucose results that relate
either to whole blood or to plasma. To see whether your blood glucose meter displays results
relating to whole blood or to plasma, see the package insert that came with your test strips.
There you will also find information on how the system works, on the test principle and on ref-
erence methods.
Measuring range 0.6–33.3 mmol/L (10–600 mg/dL)
Sample size approx. 1.5 µL (1 microlitre = 1 thousandth of a millilitre)
Measuring time approx. 5 seconds (depending on concentration)
System operating +10°C to +40°C
temperature
AC_Voicemate_Manual_EN-CAN.indd 235AC_Voicemate_Manual_EN-CAN.indd 235 1.5.2007 13:17:561.5.2007 13:17:56