For sighted persons: Using the meter and cleaning the devices
F
a
r
be
: P 2
8
1 – Pr
ü
fmi
tte
ln
u
mm
e
r: 2
05
6
24.1.2 Atmospheric humidity
Only test blood glucose at a relative humidity of between 20% and 85%.
Keep the meter and lancing device at a relative humidity of between 20% and 85%.
Sudden changes in temperature cause condensation to form in or on the meter. If this happens, do not turn the meter
on. Make sure that it returns to room temperature slowly. Never keep it in a room that is likely to harbour condensa-
tion (e.g. a bathroom).
24.1.3 Sources of interference
Strong electromagnetic fields may interfere with the proper operation of the meter. Do not use the meter close to
sources of strong electromagnetic radiation.
To avoid electrostatic discharge, do not use the meter in a very dry environment, especially one in which synthetic
materials are present.
Keep the meter away from very strong light sources (e.g. direct sunlight, studio lighting, spot lights, etc.). These
may interfere with the proper functioning of the meter and cause error messages to be displayed.
24.2 Voice unit
The voice unit may be used under the same environmental and storage conditions as the blood glucose meter.