For sighted persons: Using the meter and cleaning the devices
6
A test can be performed at the limit of the permitted range (between +6°C and +10°C or between +40°C and +44°C). In
this case the thermometer symbol is displayed.
Do not use results obtained at the limits of the permitted temperature range as a basis for making therapeutic
decisions, as they could be incorrect. Incorrect results can cause the wrong therapeutic decision to be taken and
so produce serious adverse health effects.
Tests cannot be performed at temperatures below +6°C and above +44°C.
The following message is displayed:
At temperatures below 0°C it is impossible to turn the meter on.
Keep the lancing device at temperatures between -25°C and +70°C.
Never expose the lancing device at high temperatures (e.g. in a hot car). High temperatures can cause parts of
the lancing device to become warped, so impairing its operation.
When you aren’t using the lancing device, keep it without a lancet inserted.
Never try to speed up warming or cooling of your meter, e.g. by placing it on a radiator or in a refrigerator. Doing
so can damage the meter and cause it to give incorrect results. Incorrect results can cause the wrong therapeu-
tic decision to be taken and so produce serious adverse health effects.