User`s manual
232
For sighted persons: Using the meter and cleaning the devices
F
a
r
be
: P 2
8
1 – Pr
ü
fmi
tte
ln
u
mm
e
r: 2
05
6
24 Measurement and storage conditions
The reliability of your voice unit, meter and lancing device as well as the accuracy of your results is influenced by various fac-
tors. Pay close attention to the following points.
24.1 Blood glucose meter and lancing device
24.1.1 Temperature range
Without batteries and without test strip drum the meter may be stored at between -25°C and +70°C.
With batteries and without test strip drum the meter may be stored at between -10°C and +50°C.
At temperatures above +50°C the batteries could leak and damage the meter.
At temperatures below -10°C the batteries do not have enough power to keep the internal clock functioning.
With batteries and with test strip drum the meter may be stored at between +2°C and +30°C.
For performing blood glucose tests the permitted range is from +10°C to +40°C.
AC_Voicemate_Manual_EN-CAN.indd 232AC_Voicemate_Manual_EN-CAN.indd 232 1.5.2007 13:17:541.5.2007 13:17:54