For sighted persons: Using the meter and cleaning the devices
5
This message/symbol
appears ...
and means:

The bar code on the test strip drum is damaged. To troubleshoot, see error message E-3.
A meter error has occurred.
Press the Test button to turn the meter off and then on again. If the error message per-
sists, please contact your local customer support and service centre (see Chapter 31).
During a data download both arrows fail to start flashing. The meter turns itself off automati-
cally after 5 minutes. The PC, handheld computer or the Accu-Chek Smart Printer System
also displays an error message.
A download error occurred.
Start the download again.
A strong light source is too close to the two infrared windows.
Move away from the light source and start the download again.
The infrared windows are too far apart or not properly pointing toward one another.
Reduce the distance between the two infrared ports to 10-20 cm and position them so
they are pointing towards one another. Start the download again.
