This message/symbol
appears ...
and means:
The meter has been exposed to severe differences in temperature.
The temperature is too low (between 0°C and +6°C) or too high (above +44°C) to perform
a test.
Press the Test button to turn the meter off. Allow it to cool down or warm up slowly. During
testing the permitted temperature range is from +10°C to +40°C.
You applied too little blood or control solution.
Turn the meter off and on again. Repeat the test with a new test strip and a larger amount
of blood or control solution.
There is no test strip drum in the meter.
Press the Test button to turn the meter off. Insert a new test strip drum.
The mock drum (empty drum without test strips or a label) is still in the meter.
Press the Test button to turn the meter off. Insert an Accu-Chek Compact test strip drum.
You inserted a test strip drum from a different blood glucose measuring system.
Press the Test button to turn the meter off. Insert an Accu-Chek Compact test strip drum.