For sighted persons: Using the meter and cleaning the devices
5
This message/symbol
appears ...
and means:
The batteries are empty or a meter error has occurred.
Press the Test button to turn the meter off and then on again. If the error message appears
again, the batteries may be empty. Insert new batteries. If you see the error message
again when you next turn the meter on, please contact the Accu-Chek customer care line
(see Chapter 30).
You pulled the test strip out of the meter during testing.
Press the Test button to turn the meter off and then on again. Repeat the test with a new
test strip. Only remove the test strip by pressing the Test button to turn the meter off.
The test strip is stuck.
The test strip drum does not rotate.
Press the Test button to turn the meter off. Open the drum compartment and, if
necessary, pull out the test strip. Press the red drum release button and remove the test
strip drum. Insert the test strip drum and close the drum compartment. The meter then
rotates the drum and reads the bar code. The drum symbol and bottle symbol then flash in
the display. The 90-day use-by period will no longer be properly calculated for this drum.
Perform a quality control check. You may use all the test strips remaining in the drum. The
drum symbol continues to flash in the display until you have inserted a new drum.