For sighted persons: Using the meter and cleaning the devices
F
a
r
be
: P 2
8
1 – Pr
ü
fmi
tte
ln
u
mm
e
r: 2
05
5
This message/symbol
appears ...
and means:

You turned the meter off and on again. The old test strip is still in the meter.
Hold the meter upright so the test strip is pointing downwards. Press the Test button to
turn the meter off and allow the test strip to drop out. If the test strip does not drop out,
remove it manually, even if you feel some resistance.
You pulled the test strip out of the meter before applying blood or control solution to the
strip.
Press the Test button to turn the meter off and then on again. Repeat the test with a new
test strip. Only remove the test strip by pressing the Test button to turn the meter off.
The test strip that was advanced is defective.
Hold the meter upright so the test strip is pointing downwards. Press the Test button to
turn the meter off and allow the test strip to drop out. Press the Test button to turn the
meter on again and repeat the test.
There is condensation in the meter.
Press the Test button to turn the meter off. Leave it to dry at room temperature.