User`s manual
208
For sighted persons: Using the meter and cleaning the devices
F
a
r
be
: P 2
8
1 – Pr
ü
fmi
tte
ln
u
mm
e
r: 2
05
5
This message/symbol
appears ...
and means:
The results in memory may be incorrect. Incorrect results can cause the wrong thera-
peutic decision to be taken and so produce serious adverse health effects.
Although you performed
a blood glucose test, the bot-
tle symbol is displayed.
An assignment error has occurred. The result is correct, but has been saved as a quality con-
trol test.
Press the S button once so that the bottle symbol is no longer displayed.
If the meter has already been switched off, you can no longer make the correction. You
cannot change results you have retrieved from memory with the M button.
Although you performe a
quality control check, the
bottle symbol is not dis-
played.
An assignment error has occurred. The result is correct, but has been saved as a blood glu-
cose result.
Press the S button twice so that the bottle symbol is displayed.
If the meter has already been switched off, you can no longer make the correction. You
cannot change results you have retrieved from memory with the M button.
AC_Voicemate_Manual_EN-CAN.indd 208AC_Voicemate_Manual_EN-CAN.indd 208 1.5.2007 13:17:391.5.2007 13:17:39