User`s manual

207
For sighted persons: Using the meter and cleaning the devices
5
This message/symbol
appears ...
and means:
You inserted a partly used test strip drum.
You opened the drum compartment.
You took longer than two minutes to replace the batteries (the year, date and time settings
are lost) and a partly used drum remained in the meter.
You adopted the bar code from the preceding drum, as the bar code on the drum that is
inserted is damaged (see error message E-2).
The meter cannot calculate the use-by period for this test strip drum.
You performed the requested quality control check (see explanation for the preceding display).
The 90-day use-by period for the test strip drum has expired. The test strips cannot be used.
You must not use these test strips, as they may produce incorrect results. Incorrect
results can cause the wrong therapeutic decision to be taken and so produce serious
adverse health effects.
Insert a new test strip drum.
The 90-day use-by period for the test strip drum has expired and you have performed a
quality control check (see explanation for the preceding display).
In the case of saved results: The test strips you used for this test were unusable (in the
meter for more than 90 days).
