For sighted persons: Using the meter and cleaning the devices
F
a
r
be
: P 2
8
1 – Pr
ü
fmi
tte
ln
u
mm
e
r: 2
05
5
This message/symbol
appears ...
and means:
The test strip drum in the meter is empty.
Insert a new test strip drum.
You inserted a partly used test strip drum.
You opened the drum compartment.
You took longer than two minutes to replace the batteries (the year, date and time settings
are lost) and a partly used drum remained in the meter.
You adopted the bar code from the preceding drum, as the bar code on the drum that is
inserted is damaged (see error message E-2).
You are requested to perform a quality control check in order to check the test strips. The
meter has no way of telling when the drum was first inserted or how long the partly used
drum has already been in the meter. Perform a quality control check (see Chapter 20). If the
test produces a correct result, you may continue to use the test strips. If the result is incorrect,
you need to insert a new drum. Following the quality control check, the bottle symbol disap-
pears from the display. The drum symbol continues to flash in order to draw attention to the
fact that the meter cannot calculate the use-by period. The symbol only disappears when you
insert a new drum. All results are saved along with a flashing drum symbol. If you do not per-
form a quality control check, the flashing bottle symbol appears with every test until you insert
a new drum.
AC_Voicemate_Manual_EN-CAN.indd 206AC_Voicemate_Manual_EN-CAN.indd 206 1.5.2007 13:17:381.5.2007 13:17:38