Did you perform the test within the correct temperature range (+10°C to +40°C)?
Did you use a test strip drum whose 90-day use-by period has not expired (see Chapter 6)?
Did you observe the proper storage conditions for the meter, the test strip drum and the control solution
(see Chapter 24)?
Is the control solution beyond its expiration date (see label on the bottle)?
Has the control solution been open for more than three months? Once opened, control solutions are stable for three
months. They must not be used after this period.
Did you use the correct quality control solution (Accu-Chek Compact Autocontrol)?
If you have checked these points and the results are still outside the stated concentration range, please call the Accu-Chek
customer care line (see Chapter 30).
i
Quality control results flagged with a bottle symbol are not included in statistical calculations.
i
In order for the meter to recognize the Accu-Chek Compact Autocontrol solutions automatically as control solutions
and to enable it to distinguish them from blood, the control solutions contain a dark dye. This is not poisonous. Avoid
getting the control solution on the table or your clothing. The dye is water-soluble. If you do happen to spill some, you
can wipe or rinse the dye away immediately with warm water.
i
You can dispose of empty or only partly used bottles of control solution with your regular household waste.