User`s manual
196
For sighted persons: Using the meter and cleaning the devices
F
a
r
be
: P 2
8
1 – Pr
ü
fmi
tte
ln
u
mm
e
r: 2
05
5
i
The hypo indicator cannot distinguish between blood and control solution. If you have set a hypo indicator and use
control solution G1 you may see hypo? flashing on the display. This means that the reading for the control solution is
below the blood glucose level you have set for the hypo indicator. Example: The concentration range for control solu-
tion G1 is 2.1-3.7 mmol/L (37-67 mg/dL) and the level set for the hypo indicator is 3.9 mmol/L (70 mg/dL).
Take care not to bend the test strip either before or during application of blood. Do not bend or move the test strip
while the test is in progress.
If you do bend or move the strip, it may produce an incorrect result, or you
may see the following message:
If this happens, press the Test button to turn the meter off and allow the
test strip to drop out. Repeat the quality control check with a new test
strip.
If the result is outside the stated range, perform a second quality control check. If the result of this second check is outside the
concentration range as well, please check the following points:
Did you perform the quality control check as described above?
Did you wait for the flashing test strip symbol and blood drop symbol to appear before applying control solution?
Did you wipe the tip of the dropper before you applied control solution to the surface of the washable object?
Did you compare the result with the concentration table specifically belonging to the control solution you used? Does the
concentration table belong to the test strips you are using?
AC_Voicemate_Manual_EN-CAN.indd 196AC_Voicemate_Manual_EN-CAN.indd 196 1.5.2007 13:17:311.5.2007 13:17:31