For sighted persons: Using the meter and cleaning the devices
5
If the bottle symbol is not displayed:
Press the S button twice when the result is displayed. The bottle symbol appears in the display.
i
You can only flag the result as a quality control result as long as you have not yet turned the meter off. Once the
meter has been switched off, you can no longer flag the result. You cannot change results you have retrieved from
memory with the M button.
As soon as the result is displayed and you have flagged it as a quality control test (where this is necessary):
Compare the result with the concentration table printed on the peel-off label attached to the drum container. Make sure
you are comparing the result with the correct concentration table for your control solution.
Hold the meter over a waste bin with the test strip pointing downwards.
Press the Test button to turn the meter off. The used test strip drops out.
Always let the meter eject the test strip as soon as you have compared the result with the concentration table. Do not
leave the meter lying around with the test strip still in and do not pull the test strip out of the meter. If you do, control
solution taken up by the test strip may contaminate the meter.
Wipe the tip of the dropper with a clean, dry paper towel and tightly close the bottle tightly.