For sighted persons: Using the meter and cleaning the devices
F
a
r
be
: P 2
8
1 – Pr
ü
fmi
tte
ln
u
mm
e
r: 2
05
5
To perform a quality control check you need the concentration tables for the control solution you are using. These can be found
on the peel-off label attached to the drum container. When conducting the quality control test, always check that the peel-off
label with the concentration tables belongs to the test strip drum that is currently in the meter.
20.2 Performing a quality control check
Read the package insert that came with the control solution.
Press the Test button to turn the meter on. Shortly afterwards the meter automatically advances a test strip.
Check that all the elements shown in the illustration are properly displayed on your meter.
If any of the display elements are missing, please contact the Accu-Chek customer care
line. For instance, if you can see elements other than the numerals 888 on your display
(see a possible example in the illustration right), the meter will not be able to display your
test results.
Wait until the flashing test strip symbol and blood drop symbol are displayed and the
meter beeps (provided the beep tone is turned on).
You now have about 5 minutes in which to apply control solution. After this time, the meter automatically turns off. If this
occurs, press the Test button to eject the test strip. The strip can no longer be used. Start the quality control check again with
a new test strip.