If the most recently saved result is a quality control result, this is announced in the same way as any blood glucose
result. There is no acoustic indication to signify quality control results.
19.3 Warnings and error messages
In the case of warnings, flashing symbols appear in the display. Additionally, in Acoustic Mode, the meter emits three double
beeps. Please refer to the chapters “Symbols shown in the display” and “Display messages and troubleshooting” for the com-
plete list.
Warnings may be issued at various times.
When you turn the meter on: The batteries are running low: You can now perform approximately 50 more tests.
While you are turning the meter off: You have used the last test strip. The test strip drum is empty.
Following a measurement, i.e. following the double beep that signifies that measurement is complete, and before the
result is announced:
The temperature during measurement was outside the permitted range of +10°C to +40°C.
The 90-day use-by period for the test strip drum has expired. Insert a new test strip drum.
You inserted a partly used test strip drum.
You opened and closed the drum compartment cover while there was a partly used test strip drum in the meter.
You took longer than two minutes to replace the batteries (the year, date and time settings are lost) and left a partly
used drum in the meter.