Error messages and troubleshooting
F
a
r
be
: P 2
8
1 – Pr
ü
fmi
tte
ln
u
mm
e
r: 2
05
4
Message text Cause Remedy
While the voice unit and meter are in
contact with one another
Saved error message (Menu 5.1)
Meter error E 22:
No drum in the meter, or
drum from different blood
glucose monitoring system
is inserted, or bar code is
damaged.
At the time of the error
message there was no
drum in the meter, or a
drum from a different
blood glucose monitoring
system was inserted, or
the bar code on the drum
was damaged.
i
This error message is not
announced when the voice unit
is in contact with the meter at
the time of the error message.
You can only summon the error
message in Menu 5.1.
If you have not yet eliminated the
error:
Press the Test button to turn the me-
ter off. Insert an Accu-Chek Compact
test strip drum.
If an Accu-Chek Compact test strip
drum is inserted, the bar code is
damaged. Press the Test button to
turn the meter off.
If you inserted the fi rst drum from
a pack: Remove the drum from the
meter and return it to its container.
Insert another drum from the same
pack in the meter and use all 17 test
strips. When it is time to replace the
drum, insert the fi rst drum with the
damaged bar code. When the error
message is announced, press the S
button on the meter. The meter uses
the bar code from the preceding
drum.