Error messages and troubleshooting
4
Message text Cause Remedy
While the voice unit and meter are in
contact with one another
Saved error message (Menu 5.1)
Meter error E 5: Blood was
applied too soon, or test
strip was bent or is defec-
tive, or measuring window
of meter is dirty.
You applied blood too soon
to the test strip, or bent the
test strip while applying
blood or during testing, or
the test strip is defective.
Turn the meter off and then on again
by pressing the test button. Repeat
the test with a new test strip. Wait un-
til you hear the “Apply blood” prompt.
Now (and no earlier than this) apply
blood to the test strip.
none retrospectively
Also, the measuring
window of the meter may
be dirty.
Press the Test button to turn the me-
ter off. Have the measuring window
cleaned as described in the User’s
Manual for the meter.
If you have not yet eliminated the
error:
Press the Test button to turn the me-
ter off. Have the measuring window
cleaned as described in the User’s
Manual in Chapter 22.1.