Error messages and troubleshooting
F
a
r
be
: P 2
8
1 – Pr
ü
fmi
tte
ln
u
mm
e
r: 2
05
4
Message text Cause Remedy
While the voice unit and meter are in
contact with one another
Saved error message (Menu 5.1)
Meter error E 5: Measuring
window of meter is dirty,
or you turned the meter off
and on again and there is
still an old test strip in the
meter, or there is conden-
sation in the meter.
The measuring window
may be (have been) dirty.
Press the Test button to turn the me-
ter off. Have the measuring window
cleaned as described in the User’s
Manual in Chapter 22.1.
If you have not yet eliminated the
error:
Press the Test button to turn the me-
ter off. Have the measuring window
cleaned as described in the User’s
Manual for the meter.
If there is/was an old test
strip in the meter:
Press the test button to turn the
meter off. Allow the test strip to drop
out. If the test strip does not drop out,
remove it manually, even if you feel
some resistance.
If you have not yet eliminated the
error:
Press the Test button to turn the
meter off. Allow the test strip to drop
out. If the test strip does not drop out,
remove it manually, even if you feel
some resistance.
If there is/was condensa-
tion in the meter:
Press the Test button to turn the
meter off. Leave it to dry at room
temperature.
If you have not yet eliminated the
error:
Press the Test button to turn the
meter off. Leave it to dry at room
temperature.