Operation Manual
Putting the ACCU-CHEK® Spirit Combo Insulin Pump in
Communication Mode
The built
-
in infrared interface on the bottom of the
insulin pump enables the user to
transfer data between the insulin pump and the computer.
To prepare the
insulin pump for communication:
1.
Put the
insulin pump in
STOP mode.
2.
Press
to move to the
Data Transfer
screen shown on the insulin pump's display. Press
to select.
3.
Press
.
Show me how to put the insulin pump in communication mode.
4.
Set the
insulin pump on a flat surface.
5.
Point the bottom of the insulin pump toward the front of the supported infrared cable. The insulin pump
should be approximately 4 inches (10 cm) from the infrared cable. Do not move the insulin pump or the
cable during the communication between the insulin pump and the computer. The
software is now ready to communicate information from the insulin pump.
6.
When the data transfer is complete press
to return to the
STOP mode.
Note:
The insulin pump automatically returns to the
STOP
mode 15 minutes after the last data transfer when the
Data Transfer screen in selected.
Press
and
twice simultaneously to exit while data is being transferred. The data transfer is interrupted,
and an error message occurs.