Instruction Manual

AH-TX Series Handheld Pulse Oximeter Instruction Manual
55
Alarm Setup Factory Default
SpO
2
Alarm Level High
PR Alarm Level Medium
C.1.2 Alarm Type Setup
SpO
2
Setup Adult Pediatric Neonate
SpO2 High Limit 100 100 95
SpO2 Low Limit 90 90 90
PR Setup Adult Pediatric Neonate
PR High Limit 120 160 200
PR Low Limit 50 75 100
C.3 System Setup
System Setup Factory Default
Alarm Volume 1
Beep Volume 0
Key Volume 0
Brightness 1
Scan Speed 25mm/s
C.4 Work Type Setup
Work Type Setup Factory Default
Interval 10s
H