Service manual

Service Manual Specifications 61
Nellcor
®
sensor accuracy guide
Accuracy specifications are based on controlled hypoxia studies with healthy,
non-smoking adult volunteers over the specified saturation SpO
2
range. Pulse oximeter
SpO
2
readings were compared to SaO
2
values of drawn blood samples measured by
hemoximetry. All accuracies are expressed as + “X” digits. This variation equals plus one
standard deviation (+ 1 SD), which encompasses 68% of the population.
SpO2 Accuracy (MP205 specification - 2002 and
earlier)
±3% in the range of 70-100%
Oxygen saturation (1 stand deviation).
<70% unspecified by the OEM.
Pulse Rate Range 25 to 240 bpm
Pulse Rate Range (using SpO
2
determination MP205
specification -2002 and earlier)
25 to 245 bpm
Pulse Rate Accuracy ±3 bpm
±3 bpm (low perfusion)
Pulse Rate Accuracy (MP205 Spec - 2002 and earlier) SpO
2
Module Heart Rate ±3 bpm
Table 20. OxiMax Sensor Models, Single Patient Use
Sensor Models SpO
2
Range
70% to 100%
MAX-AI ±2
MAX-PI ±2
MAX-II ±2
MAX-RI
1
± 3.5
1
The accuracy specification has been determined between saturations of 80% to 100%.
Table 21. OxiCliq Sensor Models, Single Patient Use
Sensor Models SpO
2
Range
70% to 100%
OXICLIQ-PI ± 2.5
Table 22. Reusable Sensor Models
Sensor Models SpO
2
Range
70% to 100%
D-YS (Infant to Adult) ± 3
D-YS and D-YSE ± 3.5
D-YS and D-YSPD ± 3.5
DS-100A ±3
OXI-A/N (Adult) Adult: ± 3
OXI-P/I (Pediatric/infant) ± 3