Service manual
Service Manual Disassembly and repair 49
To upgrade the pneumatic tubing:
1. Verify you have two pieces of silicon tubing 1.5 inches in length (421051-11) and one
elbow fitting (703843) available.
2. Remove the 4.4 inch silicon tubing from the âTâ fitting.
3. Insert one of the 1.5 inch silicon tubing pieces onto the âTâ fitting.
4. Insert the elbow fitting into the 1.5 inch silicon tubing.
5. Insert the second 1.5 inch tubing onto the elbow fitting.
6. Connect the 1.5 inch tubing to the Spot Vital Signs.
To assemble the pump filter:
1. Verify you have three pieces of silicon tubing .8 inches in length (421051-5), one elbow
fitting (703843), one NIBP filter (600-0520-00), and one foam sleeve (706327)
available.
2. Connect one piece of the silicon tubing at each end of the elbow fitting, and then
connect the NIBP filter to one of the open ends.
3. Connect the last piece of silicon tubing to the opposite end of the NIBP filter.
4.4 inch Silicon
Tubing