Service manual
Service Manual Disassembly and repair 41
Temperature disassembly
Follow the previous steps and then:
1. Remove the three #10 Torx screws from the temperature housing.
2. Roll the temperature housing toward the top of the Spot Vital Signs and carefully lay it
on the bench.
3. Unplug the flex cable from the connector on the temperature PCB.
4. Remove the two #10 Torx screws from the temperature PCB from the temperature
housing.
5. Remove the two #10 Torx screws on the temperature probe holder and gently pull
out.
LEAD ACID BATTERY
M
odel No
. 4200-B4
Temperature Housing
Temperature PCB
Temperature connector PCB
screws
Temperature probe holder screws