Service manual

28 Calibration Welch Allyn Spot Vital Signs
Date/time set
After recharging a dead battery or after disconnecting the battery for a few minutes, you
must program the date and time screen.
To set the date/time after after a power loss (Version 1 only):
1. Press and hold the Blood Pressure Start/Stop + Power buttons to enter the Internal
Configuration mode. The Spot Vital Signs displays an E38 error. Press the C button to
cancel the error, and the revision level of the internal software displays.
2. Press the Mode button to advance to the Date Set Screen. The day, month, and year
show in the systolic, diastolic, and heart rate displays, respectively.
3. Use the Mode button to select the date item for change. When a date option is
selected, the respective display flashes.
4. Use the Next Patient/Clear or Blood Pressure Start/Stop buttons (arrow up or
arrow down) to change the selected date option. After making all the desired date
changes, press the Mode button once to save the change and advance to the Time
Set Screen.
When in the Time Set Screen the hour (in 24-hour format) and minutes appear in the
systolic and diastolic displays, respectively. Use the Mode button to select the time
item for change. When selected, the time option flashes. Use the Next Patient/Clear
or Blood Pressure Start/Stop buttons to set the time (in the same manner as
described previously).
5. Press the Mode button to save the time and advance to the next screen.
6. Press the Power button to turn off the Spot Vital Signs.
To update the displayed date and time:
1. Press and hold the Blood Pressure Start/Stop + Power buttons to enter the Internal
Configuration Menu. Spot Vital signs displays the revision level of the internal
software.
2. Press the Mode button to advance to the Date Set Screen. The day, month, and year
appear in the systolic, diastolic, and heart rate displays, respectively.
3. Use the Mode button to select the date option for change. When a date item is
selected, the respective display flashes.
4. Use the Next Patient/Clear or Blood Pressure Start/Stop buttons (arrow up or
arrow down) to change the selected date option. After making all the desired date
changes, press the Mode button once to save the change and advance to the Time
Set Screen.
When in the Time Set Screen the hour (in 24-hour format) and minutes appear in the
systolic and diastolic displays, respectively. Use the Mode button to select the time
item for change. When selected, the time option flashes. Use the Next Patient/Clear
or Blood Pressure Start/Stop buttons to set the time (in the same manner as
described previously).
5. Press the Mode button to save the time and advance to the next screen.
6. Press the Power button to turn off the Spot Vital Signs.