Service manual

Service Manual Functional overview 23
To begin functional verification of the SureTemp thermometer:
1. Remove the probe from the probe well and clean it with either a 70% isopropyl
alcohol solution, a 10% chlorine bleach solution, or a non-staining disinfectant. Let
the probe air dry. Do not apply a probe cover.
2. Place the thermometer in Monitor Mode, refer to the thermometer’s Operator's
Manual.
3. Insert the probe into the Thermistor Device Port.
4. Wait for approximately one minute or until temperature on the thermometer is stable
for ten seconds. Compare the thermometer's temperature reading to the 9600 Plus
Calibration Set Point Temperature. If the temperatures are within ±0.1° C (±0.2° F),
the thermometer is within calibration.
5. Test all available thermometers for calibration verification at the current Calibration Set
Point Temperature. Proceed to the next Calibration Set Point Temperature, see
“Changing the calibration set point temperature” on page 23.
Changing the calibration set point temperature
To scroll from one set point to the next, press and hold the Temperature Selection button
until a beep is heard. The newly selected set point appears in the upper left corner of the
LCD display. The device’s current temperature is displayed, starts to flash, and continues
flashing until the cavity reaches the equilibrium at the new set point.
Caution Store thermometers for testing in the same room as the 9600 Plus
Calibration Tester for approximately 30 minutes prior to testing to allow for
thermal accommodation.