Service manual
2 Introduction Welch Allyn Spot Vital Signs
WARNING The Welch Allyn Spot Vital Signs is not intended for continuous
monitoring. Do not leave the device unattended while taking measurements on a
patient.
WARNING To ensure patient safety, use only accessories and supplies (i.e.,
blood pressure cuffs, hoses, temperature probes, SpO
2
sensors, etc.)
recommended for or supplied with Spot Vital Signs. Using unapproved
accessories with Spot Vital Signs can affect patient and/or operator safety.
WARNING This device is not suitable for use in the presence of a flammable
anesthetic mixture with air or oxygen or nitrous oxide. An explosion may result.
WARNING Avoid compression of the blood pressure cuff tubing or pressure
hose of the Welch Allyn Spot Vital Signs. Compression of the cuff tubing or
pressure hose may cause system errors to occur in the device.
WARNING Care should be taken to prevent water or other fluid from entering
any connectors on the device. Should this occur, the connectors should be dried
with warm air. All operating functions should then be checked for proper
operation.
WARNING Any Spot Vital Signs which has been dropped or damaged should be
checked by qualified service personnel to ensure proper operation prior to use. Do
not use the Welch Allyn Spot Vital Signs if you notice any signs of damage.
Contact the Welch Allyn Customer Service Department for assistance.
WARNING Every three months, inspect the temperature probe, SpO
2
cord, and
accessories for fraying or other damage. Replace as necessary.
WARNING There are no user-serviceable parts inside the device other than
battery replacement. Refer Spot Vital Signs to the Authorized Service Center.
WARNING The Spot Vital Signs should not be used on patients who are linked
to heart/lung machines.
WARNING The Spot Vital Signs does not operate effectively on patients who are
experiencing convulsions or tremors.
WARNING This device complies with current required standards for
electromagnetic interference and should not present problems to other
equipment or be affected by other devices. As a precaution, avoid using this
device in close proximity to other equipment.
WARNING This device is not intended for hand-held use during operation.
WARNING Welch Allyn recommends leaving the battery in the device,
regardless if the device is not used for long periods of time, since there is no
hazard of leaving the battery in the device.
WARNING Do not autoclave.
WARNING Welch Allyn is not responsible for the integrity of any mounting
installation. Welch Allyn recommends that the customer contact their Biomedical
Engineering Department or maintenance service to ensure professional
installation for safety and reliability of any mounting accessory.