Reference Manual

BiWaze Cough - User Reference Manual
62
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11
Class B
The device is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies building used for domestic
purpose.
Harmonic emissions IEC 61000-
3-2
Class A
Voltage fluctuations/Flicker
emissions
IEC 61000-3-3
Complies
13.2
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
This device is intended for use in the electromagnetic environment specified below.
The user of this device should make sure it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic
Environment -
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete
or ceramic tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Electrical fast
Transient/burs
t
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input-
output lines
±2 kV for supply mains
Not Applicable
Mains power quality should
be that of a typical home or
hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to line(s)
±1 kV line(s) to line(s)
Not Applicable
Mains power quality should
be that of a typical home or
hospital environment.