• Inhale Pressure not to exceed 85 cmH2O for 1 minute
• Exhale Pressure not to exceed -75 cmH2O for 5 secs
• Duration of inhale phase in Auto Mode within ± (10% of the setting + 0.5
seconds)
• Duration of exhale phase in Auto Mode within ± (10% of the setting + 0.5
seconds)
• All breath phases with times > 0 occurring in proper order in Auto Mode