Reference Manual

BiWaze Cough - User Reference Manual
57
ISO 10993-1 Biological evaluation of medical devices Part 1: Evaluation
and testing (Biocompatibility)
ISO 13485: 2016: Medical Devices Quality Management Systems
ISO 14971:2012 : Application of Risk Management for Medical Devices
12.5
Device classifications
AC Voltage Source
100 to 240VAC, 50/60 Hz
AC Power Supply
Input: 100-240 V, 50/60 Hz 1.0-
2.0A
IEC 60601-1 & IEC 60601-1-2
certified
Lithium Ion Battery
Power: 88 Whr
Capacity: 3200 mAh
IEC 62133-2 certified
Type of Protection Against Electric Shock
Class II
Classification According to Directive
93/42/EEC
Class IIa
Degree of Protection Against Electric Shock
Type BF Applied Part
Patient applied part
Face mask and mouth piece
Degree of Protection against Ingress
Exposure Protection, IP21
Mode of Operation:
Intermittent (30 mins ON 15
mins OFF)
12.6
Wireless communication
Feature
Dimension
Bluetooth Specification
Bluetooth Compliance
Bluetooth 4.2 Secure Connection Compliant and
CSA2 Support
Frequency
2.4 to 2.48 GHz