Input/Output Module technical data
Table 14: Input/Output Module technical data
Throughput
Minimum of 900 capped samples per hour
NOTE: The specified performance of the module is based on
measurements taken in a given test environment. The actual
performance may vary significantly, depending on the use
scenario of a track system.
Capacity
Four drawers with 125 sample tubes each for 500 sample
tubes total when a standard RackPort is used
NOTE: Based on the selected RackPort, the maximum quantity
of sample tubes can vary.
Dimensions
88 cm (width) x 103 cm (depth) x 200 cm (height)
Floor requirement
500 kg/m
2
Acoustic level
< 65 dBA
Electrical specifications:
Supply voltage
220 V to 230 V
Supply frequency
50 Hz/60 Hz
Power
Power typical 105 W
Power maximum 250 W
300 Volt-Ampere maximum
NOTE: Electrical lines are routed in the
power ducts of the GLP systems Track.
NOTE: The name plate is located inside the module and is visible when the module cover is
open. To open the module cover, the module status must be Offline.
Related information...
Technical data, page 82
Sample tube specifications, page 83
Open and close the front and rear module covers (IOM), page 303