Use Manual
Table Of Contents
- Table of contents
- Read me first
- System documentation
- Section 1: Use or function
- Section 2: Installation procedures and special requirements
- Section 3: Principles of operation
- Section 4: Performance characteristics and specifications
- Section 5: Operating instructions
- Start the GLP systems Track
- Shut down the GLP systems Track
- Perform an emergency shutdown
- Log on to the Track Sample Manager (TSM)
- Log on to the Track Workflow Manager (TWM)
- Track operation
- Track Sample Manager user interface
- Roles and permissions (TSM)
- Main menu screen element descriptions (TSM)
- Track / RE tab (TSM)
- Track / RE tab element descriptions (TSM)
- Controller tab element descriptions (TSM)
- Segment Controller screen element descriptions (TSM)
- Module Controller screen element descriptions (TSM)
- Circles tab element descriptions (TSM)
- Layout tab element descriptions (TSM)
- Statistic tab element descriptions (TSM)
- Modify statistical criteria (TSM)
- Firmware tab element descriptions (TSM)
- Routing Engine tab element descriptions (TSM)
- View the connections to TWM on TSM
- Operation mode tab element descriptions (TSM)
- Samples tab (TSM)
- Samples tab element descriptions (TSM)
- Search tab (Samples) element descriptions (TSM)
- Search result screen element descriptions (TSM)
- Sample Detail screen element descriptions (TSM)
- Sample History screen element descriptions (TSM)
- Search for samples (TSM)
- Remove a sample (TSM)
- Expert search tab (Samples) element descriptions (TSM)
- List requested tab (Samples) element descriptions (TSM)
- Request to send a sample to an output area (TSM)
- Buffered samples list tab (Samples) element descriptions (TSM)
- CARs tab (TSM)
- Track Workflow Manager user interface
- Roles and permissions (TWM)
- Main menu screen element descriptions (TWM)
- Menu overview screen element descriptions (TWM)
- Sample search screen (TWM)
- Messages screen (TWM)
- Instrument status screen (TWM)
- Test status screen (TWM)
- Test status screen element descriptions (TWM)
- View the test status (TWM)
- Disable the tests by instrument on the Test status screen (TWM)
- Enable the tests by instrument on the Test status screen (TWM)
- Disable the tests by type of analysis on the Test status screen (TWM)
- Enable the tests by type of analysis on the Test status screen (TWM)
- Connections screen (TWM)
- Input/Output Module operation
- Power on the module (IOM)
- Power off the module (IOM)
- Place the module online (IOM)
- Place the module offline (IOM)
- Pause the module (IOM)
- Deactivate pause mode (IOM)
- Access the Login screen (IOM)
- Access the Information screen (IOM)
- Access the Configuration screen (IOM)
- Main menu screen (IOM)
- Main menu screen element descriptions (IOM)
- Load samples into FlexRacks (IOM)
- Unload samples from FlexRacks (IOM)
- Section 6: Calibration procedures
- Section 7: Operational precautions and limitations
- Section 8: Hazards
- Section 9: Service, maintenance, and diagnostics
- Section 10: Troubleshooting
- Error messages
- Communication and software update error messages (0-10000)
- Track element error messages (10001-19999)
- Module component error messages (20000-29999)
- Message code: 20105
- Message code: 20150
- Message code: 20151
- Message code: 20200
- Message code: 20201
- Message code: 20202
- Message code: 20203
- Message code: 20204
- Message code: 20205
- Message code: 20206
- Message code: 20207
- Message code: 20209
- Message code: 20210
- Message code: 20211
- Message code: 20300
- Message code: 20900
- Message code: 20901
- Message code: 20902
- Message code: 20903
- Message code: 20904
- Message code: 20905
- Message code: 20906
- Message code: 20907
- Message code: 20908
- Message code: 20909
- Message code: 20910
- Message code: 20911
- Message code: 20912
- Robot picola error messages (21000-21060)
- Message code: 21000
- Message code: 21001
- Message code: 21002
- Message code: 21003
- Message code: 21004
- Message code: 21005
- Message code: 21006
- Message code: 21008
- Message code: 21009
- Message code: 21010
- Message code: 21011
- Message code: 21012
- Message code: 21013
- Message code: 21014
- Message code: 21015
- Message code: 21016
- Message code: 21017
- Message code: 21018
- Message code: 21019
- Message code: 21021
- Message code: 21023
- Message code: 21034
- Message code: 21036
- Message code: 21037
- Message code: 21041
- Message code: 21050
- Message code: 21060
- Archive error messages (21300-21468)
- Centrifuge error messages (21800-22018)
- Display error messages (23000-23017)
- Message code: 23100
- Drawer error messages (25000-25107)
- Message code: 25000
- Message code: 25001
- Message code: 25002
- Message code: 25003
- Message code: 25005
- Message code: 25006
- Message code: 25009
- Message code: 25010
- Message code: 25014
- Message code: 25015
- Message code: 25016
- Message code: 25017
- Message code: 25018
- Message code: 25019
- Message code: 25028
- Message code: 25032
- Message code: 25100
- Message code: 25101
- Message code: 25102
- Message code: 25103
- Message code: 25104
- Message code: 25105
- Message code: 25150
- Message code: 27000
- Laboratory automation system error messages
- Track Sample Manager errors
- Track element errors (700-16000)
- Controller interfaces sent by module controllers (28000-30100)
- TSM generated message error messages (70000-71999)
- Message code: 70001
- Message code: 70002
- Message code: 70003
- Message code: 70004
- Message code: 70005
- Message code: 70006
- Message code: 70007
- Message code: 70008
- Message code: 70009
- Message code: 70011
- Message code: 70015
- Message code: 70016
- Message code: 70017
- Message code: 70018
- Message code: 70019
- Message code: 70021
- Message code: 70022
- Message code: 70023
- Message code: 70026
- Message code: 70029
- Message code: 70031
- Message code: 70032
- Message code: 70033
- Message code: 70034
- Message code: 70035
- Message code: 70036
- Message code: 70037
- Message code: 70038
- Message code: 70039
- Message code: 70040
- Message code: 70041
- Message code: 70042
- Message code: 70043
- Message code: 70044
- Message code: 70045
- Message code: 70047
- Message code: 70048
- Message code: 70050
- Message code: 70051
- Message code: 70052
- Message code: 70053
- Message code: 70054
- Message code: 70062
- Message code: 70063
- Message code: 70074
- Message code: 70075
- Message code: 70076
- Message code: 70077
- Message code: 70078
- Message code: 70079
- Message code: 70080
- Message code: 70086
- Message code: 70087
- Message code: 70088
- Message code: 70089
- Message code: 70090
- Message code: 70091
- Message code: 70094
- Message code: 70096
- Message code: 70097
- Message code: 70098
- Message code: 70099
- Message code: 70100
- Message code: 70101
- Message code: 70102
- Message code: 70103
- Message code: 70104
- Message code: 70105
- Message code: 70107
- Message code: 70108
- Message code: 70109
- Message code: 70112
- Message code: 70113
- Message code: 70114
- Message code: 70115
- Message code: 70116
- Message code: 70117
- Message code: 70118
- Message code: 70119
- Message code: 70120
- Message code: 70121
- Message code: 70122
- Message code: 70123
- Message code: 70124
- Message code: 70125
- Message code: 70126
- Message code: 71000
- Message code: 71001
- Message code: 71002
- Message code: 71003
- Message code: 71004
- Message code: 71005
- Message code: 71006
- Message code: 71007
- Message code: 71008
- Message code: 71009
- Message code: 71010
- Message code: 71011
- Message code: 71012
- Message code: 71013
- Message code: 71014
- Message code: 71015
- Message code: 71016
- Message code: 71017
- Message code: 71018
- Warning messages of controller interfaces sent by analyzers (80000-80999)
- Message code: 80000
- Message code: 80001
- Message code: 80002
- Message code: 80003
- Message code: 80004
- Message code: 80005
- Message code: 80006
- Message code: 81011
- Message code: 80007
- Message code: 81005
- Message code: 81006
- Message code: 81010
- Message code: 81012
- Message code: 81013
- Message code: 81014
- Message code: 81015
- Message code: 81016
- Message code: 81017
- Message code: 81019
- Message code: 81020
- Message code: 81022
- Message code: 81023
- Message code: 81027
- Message code: 81028
- Message code: 81029
- TWM generated message error messages (81000-82999)
- Observed problems
- CAR observed problems
- New CAR does not move
- Sample tube is slanted in the sample holder
- CAR position is skewed on the track
- CAR is caught on the charge lane charging pin or lane element transition
- CAR movement is jerky or noisy
- Rechargeable lithium battery is not charged
- Rechargeable lithium battery is overheating
- CAR lurches from side to side and is not stopped at the AccessPoint
- CAR LEDs blink in an uncontrolled manner (such as red and yellow) while moving
- CAR always moves at a constant speed
- CAR does not exit the charge lane
- CARs come into contact with the charge lane
- CARs do not move
- CARs are present in the charge lane but cannot move
- CARs move too slowly
- CAR does not maintain distance with the CAR ahead of it or pushes it forward
- Track observed problems
- Track Sample Manager observed problems
- Input/Output Module observed problems
- CAR does not move to the module
- Error message is displayed
- CAR is not gripped at the AccessPoint
- CAR is not permanently held in place at the AccessPoint
- Robot does not work
- Gripper loses its grip on the sample tube
- Sample tube is jammed in the gripper
- Sample tube is tilted in the FlexRack
- Bar code is not read
- CAR observed problems
- Miscellaneous corrective action procedures
- Error messages
- Revision history
- Glossary
- accessories
- AccessPoint
- analyzer
- bar code label
- biological hazard
- CAR
- circle
- Clinical and Laboratory Standards Institute
- CLSI
- diagnostic procedure
- discipline
- FlexRack
- GLP systems Track
- Input/Output Module
- instrument status
- IOM
- laboratory automation system
- laboratory information system
- lane element
- LAS
- LIS
- loading area
- logon (TSM)
- maintenance procedure
- message
- middleware
- module
- module status
- monitor
- password
- pipettor
- printed circuit board
- RackPort
- radio-frequency identification (RFID) tag
- robot
- screen
- system log
- time-out
- track
- Track Sample Manager
- track section
- Track Workflow Manager
- TSM
- TWM
- user name
- window
- Index
Sample bar code label requirements
Table 6: Sample bar code label requirements
Bar code and label placement
• Approved: Only one bar code per tube (excluding sample
labels)
IMPORTANT: Incorrect bar code label placement or
attachment on sample tubes may cause material damage
when CARs are loaded and unloaded. For example, the
sample tube may remain stuck in the gripper or the CAR
may be lifted together with the sample. Observe the
appropriate bar code label placement or attachment on
the sample tubes.
NOTE: Bar code labels should meet the recommended
guidelines and should be correctly affixed to sample
tubes in accordance with CLSI document AUTO2-
A2: Laboratory Automation: Bar Codes for Specimen
Container Identification.
• Not approved: More than one bar code per tube
(excluding sample labels)
Label stock
• Approved: Matte paper
• Not approved: Gloss or coated paper
Symbology
The following symbologies can be used on the Input/Output
Module:
• Codabar
• Code 39
• Code 128
• Interleaved 2 of 5
Related information...
Technical data, page 82
Sample tube specifications, page 83
Sample processing specifications, page 86
Bar code reader (IOM), page 57
Load samples into FlexRacks (IOM), page 168
Sample tube specifications
Table 7: Sample tube specifications
Tube type
• Approved: Rounded bottom
• Not approved:
Section 4
Performance characteristics and specifications
Technical data
GLP systems Track Operations Manual
80003968-101 - 2021-09-16
83