Use Manual

Table Of Contents
Technical data
Technical documents for a customized GLP systems Track are distributed to the owner in
separate attachments. The following separate documents are available:
Track layout
Energy plan
Service documentation
Supplier documentation (optional)
Related information...
Performance characteristics and specifications, page 81
Environmental specifications and requirements, page 82
Sample bar code label requirements, page 83
Sample tube specifications, page 83
Sample processing specifications, page 86
Track electrical specifications, page 87
CAR technical data, page 87
Track Sample Manager technical data, page 88
Track Workflow Manager technical data, page 88
Input/Output Module technical data, page 89
Environmental specifications and requirements
Table 5: Environmental specifications and requirements
Altitude
30.8 m (100 ft) below sea level to 2000 m (6561 ft)
above sea level
Relative humidity during operation
30% to 80%, noncondensing
Acoustic level
< 65 dBA
Placement
Install indoors only.
Storage and transport
Keep dry.
Fragile: Handle with care.
Ambient temperature during operation
15°C to 30°C
Related information...
Technical data, page 82
Performance characteristics and specifications
Technical data
Section 4
82
GLP systems Track Operations Manual
80003968-101 - 2021-09-16