Bar code is not read
Probable cause Corrective action
The bar code reader is dirty or defective. 1. Clean the bar code reader. Perform Clean the bar
code readers (IOM), page 204.
2. Contact an Abbott Laboratories representative or an
authorized service representative if necessary.
The bar code is illegible. 1. Relabel the sample tube.
2. Reintroduce the sample tube to the module.
Related information...
Input/Output Module observed problems, page 296