Use Manual

Table Of Contents
Sample in CAR was exchanged, page 292
Errors at cross switch controllers, page 292
Errors at AccessPoint, page 293
Errors at charge lane, page 293
Sample is missing
Probable cause Corrective action
A sample is missing. Use the Track Sample Manager to display the last
action of the missing sample.
Related information...
Track observed problems, page 291
Operator manually loads or unloads CAR
Probable cause Corrective action
The operator removes a sample from a CAR manually or
loads a CAR manually.
CAUTION: Do not remove samples from a
CAR or the track. All samples removed from
the track must be placed back in the Input/
Output Module for appropriate routing. When
samples are removed from the track, they must
be deleted from the Track Sample Manager
before they are reloaded.
Observe that the CAR moves to the output module for
inspection purposes. Confirm that checks are performed
on the CAR loading status or on the sample.
Related information...
Track observed problems, page 291
Sample in CAR was exchanged
Probable cause Corrective action
The sample in the CAR was exchanged. Verify if there are differences in assignment between the
CAR ID and the sample. If differences in assignment are
present, the sample is transported to the error area of
the output module.
Related information...
Track observed problems, page 291
Errors at cross switch controllers
Probable cause
Corrective action
Errors occurred at the cross switch controllers. Contact an Abbott Laboratories representative or an
authorized service representative.
Troubleshooting
Observed problems
Section 10
292
GLP systems Track Operations Manual
80003968-101 - 2021-09-16