Use Manual

Table Of Contents
CAR is caught on the charge lane charging pin or lane element transition
Probable cause Corrective action
The distance between the CAR underseal and the lane
elements is insufficient.
Inspect the bottom of the CAR for dust. If
replacement of the front underseal is necessary,
perform Replace the front underseal of a CAR, page
300.
Inspect the lane elements. If replacement of
any lane elements is necessary, contact an
Abbott Laboratories representative or an authorized
service representative to replace the appropriate
lane elements.
Related information...
CAR observed problems, page 286
CAR movement is jerky or noisy
Probable cause Corrective action
The CAR is dirty. Clean the CAR. Perform one or more of the
following procedures:
Clean the sensors, page 195.
Clean the housing, page 196.
Clean the charging contacts, page 196.
Clean the sample holder, page 197.
Clean the drive wheel and the wheel arch,
page 197.
Contact an Abbott Laboratories representative or an
authorized service representative if necessary.
Related information...
CAR observed problems, page 286
Rechargeable lithium battery is not charged
Probable cause Corrective action
The charging contacts on the rechargeable lithium
battery are dirty.
Clean the charging contacts on the rechargeable
lithium battery.
Contact an Abbott Laboratories representative or
an authorized service representative to replace the
rechargeable lithium battery.
Related information...
CAR observed problems, page 286
Rechargeable lithium battery is overheating
Troubleshooting
Observed problems
Section 10
288
GLP systems Track Operations Manual
80003968-101 - 2021-09-16