TSM generated message error messages (70000-71999), page 251
Message code: 70101
Module ... does not belong to the same discipline as the other modules of the following areas: .... For
output, archive, centrifuge and analyzer areas, all modules of that area must be assigned to the same
discipline.
Probable cause Corrective action
An error has occurred. Contact Abbott service personnel
Related information...
TSM generated message error messages (70000-71999), page 251
Message code: 70102
Area ... at module ... cannot be approached. Reason: Cap status ... not supported. Supported: ...
Probable cause Corrective action
An error has occurred. Locate sample and check cap status. If not resolved,
contact Abbott service personnel
Related information...
TSM generated message error messages (70000-71999), page 251
Message code: 70103
Target ... cannot be approached. Reason: Sample tube type is unknown. It is required at module ....
Probable cause Corrective action
An error has occurred. Contact Abbott service personnel
Related information...
TSM generated message error messages (70000-71999), page 251
Message code: 70104
Target ... cannot be approached. Reason: No bottle type defined for module .... Bottle type for the
combination tube type ... and product code ... not found.
Probable cause Corrective action
An error has occurred. Contact Abbott service personnel
Related information...
TSM generated message error messages (70000-71999), page 251
Message code: 70105
Aliquot #... cannot be created. Reason: Tube type not found.