Use Manual

Table Of Contents
Clean the gripper (IOM)
Required materials
Laboratory-grade surface disinfectant
Lint-free cloth
Required module
status
Offline
Perform this as-needed procedure to clean the gripper.
1. Ensure that all samples are removed from the module.
2. Open the front module cover.
3. Dampen a lint-free cloth with a surface disinfectant.
4. Carefully wipe the gripper to remove any dust.
5. Close the front module cover.
Related information...
As-needed cleaning procedures (IOM), page 201
Clean the RackPorts (IOM)
Required materials
Laboratory-grade surface disinfectant
Lint-free cloth
Required module
status
Offline
Perform this as-needed procedure to clean the RackPorts.
1. Ensure that all samples are removed from the module.
2. Press the drawer push button to open the drawer.
CAUTION: A minimum clearance between the module and the operator is required
to avoid injury when opening the drawer.
3. Dampen a lint-free cloth with a surface disinfectant.
4. Carefully wipe the RackPorts to remove any dust.
5. Press the drawer push button to close the drawer.
Related information...
As-needed cleaning procedures (IOM), page 201
Service, maintenance, and diagnostics
Input/Output Module maintenance
Section 9
202
GLP systems Track Operations Manual
80003968-101 - 2021-09-16